IAS-TW10 Tritium Surface Contamination Wipe Monitor — No P-10 Gas
A portable counter for tritium wipe (smear) samples that needs no P-10 counting gas. Wipe a 100 cm² area, place the wipe in the instrument and read the removable tritium contamination in under 2 minutes.
Quick answer
IAS-TW10 is a surface contamination / wipe monitor in the IAS-T tritium monitoring series from Iodine Air Systems. It counts 100 cm² tritium wipe samples in under 2 minutes without P-10 gas. Portable; calibration certificate and certificate of conformance included; price on quotation.
What the IAS-TW10 does
Tritium surface contamination cannot be found with a standard frisker, because tritium beta particles (maximum 18.6 keV) do not get through a detector window. The usual method is a wipe (smear) counted by liquid scintillation, which means cocktail, vials, a lab counter and a wait.
The IAS-TW10 gives a field result in under 2 minutes and does not use P-10 counting gas, so there are no gas bottles to carry or change. It measures a 100 cm² wipe, which is the same area as the US dpm/100 cm² convention. Readings are shown in kDPM/dm² (1 kDPM/dm² = 1,000 dpm/100 cm²), nCi/dm² or Bq/cm², with an alarm level that can be set anywhere in the range.
Key features
- No P-10 (argon-methane) counting gas required
- Result in under 2 minutes
- 100 cm² measurement area, matching the dpm/100 cm² convention
- Units: kDPM/dm², nCi/dm², Bq/cm²
- Alarm level settable across the full range
- LCD touchscreen
- Portable: under 2.7 kg (6.0 lb)
IAS-TW10 specifications
| Measured sample | Wipe (smear) samples from surfaces |
| Measurement range | 0 – 1 Ci/dm² |
| Measurement area | 100 cm² (1 dm²) |
| Response time | < 2 min |
| Accuracy | ±4.44 kDPM/dm² (0–100 kDPM/dm²); ±10% (> 100 kDPM/dm²) |
| Display units | kDPM/dm² (= 1,000 dpm/100 cm²), nCi/dm², Bq/cm² |
| Alarm | Settable over the full range |
| Display / control | LCD touchscreen |
| Counting gas | None (no P-10) |
| Operating environment | 15–50 °C (59–122 °F); 5–65% RH |
| Dimensions | 150 × 178 × 254 mm (5.9 × 7.0 × 10.0 in) |
| Weight | < 2.7 kg (6.0 lb) |
| Power | Stated on quotation |
Specifications are nominal manufacturer values and may change. Confirmed values are stated on the quotation and data sheet. DAC multiples use the tritiated-water DAC of 2 × 10⁻⁵ µCi/mL (7.4 × 10⁵ Bq/m³) from 10 CFR 20 Appendix B and 10 CFR 835 Appendix A.
Typical applications
- Screening tools and equipment leaving tritium areas (confirm near limits by LSC)
- Triage surveys after spills or suspected high-level contamination
- Glovebox port and bag-out surface checks
- Tritium light and luminous-sign manufacturing lines
- Decommissioning and waste-packaging surveys
US supply, documentation and quality
Iodine Air Systems supplies the IAS-T series as an authorized distributor. Instruments are built by our manufacturing partner to our US configuration. Each quotation states the exact deliverables, lead time and country of origin.
- US power version: 120 V AC / 60 Hz for mains-powered models (export versions for 220–230 V / 50 Hz).
- Factory calibration certificate with every instrument, stating the calibration method and reference standards.
- Certificate of conformance and product certification documents supplied with the quotation.
- Specifications in SI and US units, with ranges stated in µCi/mL and in multiples of the tritium DAC (10 CFR 20 Appendix B / 10 CFR 835 Appendix A), so they drop straight into your procedures.
- Commercial-grade supply. Instruments are supplied as commercial-grade items, not as 10 CFR 50 Appendix B / NQA-1 safety-related items. For safety-related use, we support the purchaser's commercial-grade dedication with critical-characteristic data and test records.
- Price on quotation. Price depends on model, options, quantity, documentation package and destination.
Compare with related IAS-T models
Portable Tritium-in-Air Monitor
Portable survey monitor
View IAS-TP20 →IAS-TA30Floor-Standing Tritium Air Monitor
Fixed area monitor (cabinet)
View IAS-TA30 →IAS-TG30Glovebox & Process Tritium Monitor
Process monitor (sealed, all-metal)
View IAS-TG30 →See the full IAS-T tritium monitor range or use the tritium DAC converter to check a reading against your limits.
IAS-TW10 FAQ
Can a normal frisker detect tritium contamination?
No. Tritium betas have a maximum energy of 18.6 keV and cannot penetrate a typical detector window, so surface tritium is measured by wiping the surface and counting the wipe, either with liquid scintillation or a wipe monitor such as the IAS-TW10.
Does the IAS-TW10 need P-10 gas?
No. It does not use P-10 counting gas, so there are no gas cylinders to manage.
Can the IAS-TW10 replace liquid scintillation counting?
No. Use it for fast screening of higher-level contamination. Its stated accuracy at low levels is ±4,440 dpm/100 cm², so it cannot resolve common 1,000 dpm/100 cm² limits, and results near the DOE value of 10,000 dpm/100 cm² should be confirmed by liquid scintillation counting.
What units does it report?
kDPM/dm², nCi/dm² and Bq/cm². Because its measurement area is 100 cm² (1 dm²), 1 kDPM/dm² equals 1,000 dpm/100 cm².
How much does the IAS-TW10 cost?
The IAS-TW10 is priced on quotation. The price depends on options, quantity, power version, documentation package and delivery destination. Send the quote form on this page or email sales@iodineairsystems.com, and we reply with price, lead time and the list of documents supplied.
Is the IAS-TW10 available for US 120 V power?
Yes. Mains-powered IAS-T models are available in a 120 V AC / 60 Hz US version, and in 220–230 V / 50 Hz for export.
What documents come with the IAS-TW10?
A factory calibration certificate, a certificate of conformance, product certification documents and an English operating manual. A commercial-grade dedication support package is available for purchasers who dedicate the instrument for safety-related use.
Request a quote: IAS-TW10
Fill in what you know and press Email or WhatsApp. Your message is prepared for you to send. Nothing is stored on this page.
Or call +1 (307) 533-0268 · sales@iodineairsystems.com
Related pages
- How to choose a tritium monitor
- Tritium monitoring for fusion facilities
- Tritium DAC and unit converter (Bq/m³ ↔ µCi/mL)
- Tritium monitor selection guide (PDF)
- Nuclear-grade radioiodine carbon and consumables
- Cyclohexane methyl-iodide alternative test system
- Methyl iodide carbon testing service
- Nuclear air filtration systems
- Nuclear decommissioning air filtration